Hallowell Veterinary Anesthesia Ventilator Model 2002 Pro
$ 660
- Description
- Size Guide
Description
New Features in ALL 2002PRO
Ventilators
A Low Supply Gas Pressure Alarm
, now standard on all units, will sound when the supply gas pressure falls below 35 psi, 2.4bar and stops around 42 psi. The normal supply gas operating pressure is 50psi or 3.5bar. It is not until 35psi or 2.4bar that the ventilators delivered volume will begin to drop off proportionally with the waning supply pressure.
The Model 2002 became our best-selling ventilator with the addition of a fine volume control valve. Now with the Model 2002
PRO
we have simplified the Model 2002 with the replacement of the two volume control valve knobs to
one volume control valve knob
which has the functionality of both.
The single valve turns a total of nine and a half times
; the first two and a half turns utilizing the slight, minute changes for pediatric
animal
patients, and the remainder of the rotations for use with the adult and extra large bellows and housings.
All other parts and accessories for the Model 2002
PRO
have remained the same.
The Hallowell EMC Model 2002
PRO
is electronically controlled, time-cycled, pressure limited ventilator.
Electronic control
-- Eliminates the substantial driving gas required to operate pneumatically controlled ventilators.
Time cycled
-- Ensures consistently spaced breaths. The possibility of stacking breaths or auto-PEEP is eliminated by keeping the I:E ratio constant at 1:2. This means that regardless of the respiratory rate setting two-thirds of the breath will remain for expiration. The respiratory rate is adjustable from 6 - 40 breaths per minute (BPM).
Volume constant
-- Ensures that each delivered breath will have essentially the same volume, independent of changes in
animal
patient compliance and airway resistance. The deliverable tidal volume (TV) range is from approximately 20 cc to 3,000 cc.
Pressure Limited
-- Provides for
animal
patient safety by limiting the maximum working pressure (MWPL) to a user adjustable pressure range of 10 - 60 cm H
2
O.
Ventilators are equipped with dual airway pressure alarms that monitor and sound when the peak inspiratory pressure (PIP) of any breath either exceeds the preset maximum working pressure limit (MWPL); or fails to reach 6 cm H2O as in the case of a disconnect.
This version of our ventilator also includes the 0-300mL Pediatric Attachment. Included in the attachment are the housing, bellows, and adapter that sits on the base to accept the bellows.